SITRANS F C MASSFLO

2.7 Approvals

Notified body
2.8 Special conditions for
safe use
10
2. Installation
Sensor MASS 2100 DI 1.5, DI 3, DI 6, DI 15, DI 25 and DI 40
0539
II 1 G
DEMKO 03 ATEX 135252X
EEx ia IIC T3...T6
Temperature classes for ambient temperature between − − − − − 20° ° ° ° ° C to +50° ° ° ° ° C:
T3 (max. surface < 200°C) for liquid temperature lower than 180°C
T4 (max. surface < 135°C) for liquid temperature lower than 135°C
T5 (max. surface < 100°C) for liquid temperature lower than 100°C
T6 (max. surface < 85°C) for liquid temperature lower than 85°C
Signal converter MASS 6000
0539
II (1) G
DEMKO 03 ATEX 135251X
[EEx ia] IIC
Ambient temperature −20°C to +50°C
The SITRANS F C MASSFLO
UL International DEMKO A/S
Lyskaer 8, P.O. Box 514
DK-2730 Herlev
Denmark
The intrinsic safety output terminals are located in the same room as the input terminals and the
final installation shall secure the spacing of 50 mm between the intrinsic safety terminals and the
non-intrinsic safety terminals. To prevent mixing of cables between intrinsically safe terminals an
all other terminals, a isolating plate is mounted in the terminals room.
The signal converter shall always be connected to ground using the internal terminal.
The code number FDK:083H0202, FDK:083H0203, FDK:083H206 and FDK:083H0207 shall be
installed in a 19" rack with an ingress protection of at least IP 20.
For IP 65 Siemens Flow Instruments code number FDK:083F5037 can be used.
On the connection board are terminals marked „No wiring connections". These terminals are not
for field wiring but only for connection to MASS 6000 SENSORPROM
The 24 V input voltage for the 24 Volt version shall be from a safety-isolated transformer according
to EN 60742 or EN 61558-1.
SFIDK.PS.028.R1.02