2.0.2 Model CG60DL keypad
The ON/OFF key switches the gauge on or off.
To switch the gauge on press the ON/OFF key. The gauge will perform a brief test by illuminating all of the
segments in the display. After one second, the display will show the internal software version number, the
current batch location and location status, and then the measurement mode 'P-E' or 'E-E'. The display then
shows '0.00' (or '0.000' if using imperial units), indicating the gauge is ready to take readings.
To switch the gauge off press the ON/OFF key. The gauge retains all of its settings even when the power is
off. The gauge also features an 'auto-power down' function designed to conserve battery life. If the gauge is
idle for 5 minutes, it will switch itself off.
The PRB-0 key sets the gauge to zero in much the same way that a mechanical micrometer is set to zero.
PRB
If the gauge is not set to zero correctly, all of the measurements that the gauge makes may be in error by
0
some fixed value.
See "Setting up the gauge" on page 15.
When the Model CG60DL is taking readings, press the CAL key to adjust calibration.
Calibration adjustment sets the sound-velocity value that the gauge uses when calculating thickness. The
gauge will either calculate the sound-velocity from a sample of the material being measured, or allow a known
velocity value to be entered directly.
See "Calibration" on page 17.
When the Model CG60DL is taking readings, press the MODE key to adjust the features and settings of the
MODE
gauge (alarm, scan, units, P-E/E-E, backlight, and beeper). The MODE key is used in conjunction with the
arrow and SEND keys to enable/disable the features and settings.