8.2.2 To set alarm value and switch alarm on
Elcometer CG60
1. Press ON/OFF key to switch on the gauge.
2. Press ALRM key to toggle the status of the
alarm until the gauge displays:
ALAr followed by a thickness value and
flashing MM (or IN) symbol.
3. Press UP and DOWN arrow keys to adjust
thickness value.
4. Press ALRM key when correct value is
displayed.
The alarm value is now set and the alarm is switched on.
8.2.3 To switch alarm off
Repeat the steps in 8.2.2, but select ALAr OFF.
26
Elcometer CG60DL
1. Press ON/OFF key to switch on the gauge.
2. Press MODE key to activate features and
settings.
3. Press UP or DOWN arrow keys to scroll to
ALAr. The gauge will display:
ALAr OFF
4. Press SEND key. The gauge will display:
A thickness value and flashing MM (or IN)
symbol.
5. Press UP and DOWN arrow keys to adjust
thickness value.
6. Press SEND key when correct value is
displayed.
7. Press MODE key.