Flir Gf3Xx Series Series General Instrument Check - FLIR GF300 User Manual

Gf3 series
Hide thumbs Also See for GF300:
Table of Contents

Advertisement

10
FLIR GF3xx series series general
instrument check
The following general instrument check process ensures that the camera can detect the
intended gas compounds with the same sensitivity as when originally manufactured.
1. Make sure that the camera powers on.
2. Make sure that the camera completes the cool-down process and produces a live in-
frared image.
3. Make sure that the camera does not report any error messages on startup.
4. Make sure that the camera focuses properly.
5. Make sure that the camera is able to engage HSM mode.