Extending Battery Operation Time

Charging the batteries when they are not completely empty, reduces the battery
operating time for the Fluke 43. To keep the batteries in optimal condition,
observe the following guidelines:
Operate the Fluke 43 on batteries until the
the screen. This indicates that the batteries are very low and that they need to
be recharged.
To extend battery operation time, you can refresh the batteries.
During a battery refresh, the batteries will be discharged and charged
completely. A complete refresh cycle takes about 12 hours and should be
done about four times a year.
Open the main menu.
1
K
2
INSTRUMENT SETUP
K
3
START BATTERY REFRESH
Make sure the Fluke 43 is powered with the
power adapter.
4
Press
YES
Do not disconnect the power adapter.
This will interrupt the refresh cycle.
After start of the battery refresh cycle, the screen will be blank. The
backlight is on during the refresh cycle.
to continue.
Note
Maintenance
Batteries
symbol appears at the top of
2
25