10. HOT AIR GENER ATOR
Guidance and Manufacturer's Declaration – Electromagnetic Immunity
The VOLCANO MEDIC Vaporization System is intended for use in the electro-
magnetic environment specified below. The customer or the user of the VOL-
CANO MEDIC Vaporization System should assure that it is used in such an en-
vironment.
Immunity test
IEC 60601
test level
Conducted RF
3 Vrms
IEC 61000-4-6
150 kHz to 80
MHz
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz to 2,5
GHz
Compliance
Electromagnetic
level
environment – guidance
Portable and mobile RF
communications
equipment should be
used no closer to any part
of the VOLCANO MEDIC
Vaporization System,
including cables, than the
recommended separation
distance calculated from
the equation applicable to
the frequency of the
transmitter.
Recommended
separation distance :
d = 1,2 √P
3 Vrms
d = 1,2 √P
3 V/m
für 80 MHz bis 800 MHz
d = 2,3 √P
für 800 MHz bis 2,5 GHz
52
10. HOT AIR GENER ATOR
Immunity test
IEC 60601
test level
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures,
objects and people.
Compliance
Electromagnetic
level
environment – guidance
Where P is the maximum
output power rating of the
transmitter in watts [W]
according to the
transmitter manufacturer
and d is the recommend-
ed separation distance in
meters [m].
Field strengths from fixed
RF transmitters, as
determined by an
electromagnetic site
survey¹), should be less
than the compliance level
in each frequency
range²).
Interference may occur in
the vicinity of equipment
marked with the following
symbol:
53