Hot Air Generator; Cleaning; Care And Maintenance - Vapormed VOLCANO MEDIC Instructions For Use Manual

Table of Contents

Advertisement

9. HYGIENE
Cap Ring
Cap Housing
Cap Cylinder
Upper Screen
The upper Screen has to be placed
carefully into the intended notch. Non-
observance may result in plant parti-
cles in the Valve Balloon that might be
inhaled subsequently.
Assembling the Filling Chamber
Push the Filling Chamber Cylinder
from above into the Filling Chamber
Housing and allow it to lock in place.
Insert the Filling Chamber Clips in the
openings provided for them in the foot
of the Filling Chamber Housing.
Please ensure that the Filling Cham-
ber Clips are placed properly as
shown in the illustration.
Filling Chamber
Cylinder
Filling
Chamber
Housing
Filling
Chamber
Clip
Insert Filling Chamber Clips
Insert the Screen or the Liquid Pad
into the Filling Chamber from above.
Screw Filling Chamber Cap on Filling
Chamber.
44
10. HOT AIR GENER ATOR

10.1. Cleaning

Before maintenance, the device has
to be switched off and the power plug
has to be unplugged.
For home use it is sufficient to clean
the housing of the Hot Air Generator
with a damp cloth. Do not hold the de-
vice under running water.
In a physician's office or clinic, the
housing of the Hot Air Generator can
be cleaned and disinfected with a sur-
face disinfectant.
The surface disinfectant should be
certified (e.g. VAH/DGHM or FDA
approval or CE marking), be suitable
for disinfecting the Hot Air Generator,
and compatible with the Hot Air Gen-
erator (see Chapter 9.2. "Material Re-
sistance", page 38).
Use only a soft clean cloth for wiping.
The cloth may be damp, but not wet.
Prevent liquid from entering the Hot
Air Generator.
The tests for the material resistance
of the Hot Air Generator for surface
disinfection were made using the dis-
infectant "neoform MED Spray" from
Dr. Weigert.
Liquid entering the Hot Air Generator
can damage the electrical parts and
cause malfunctions.
If any liquid does get into the Hot Air
Generator, contact our Service Cent-
er immediately.

10.2. Care and Maintenance

Before any maintenance work is done,
the device has to be switched off and
the power plug has to be unplugged.
At the bottom of the Hot Air Generator
is an Air Filter.
The Air Filter must be inspected for
contamination every four weeks by
turning the Air Filter Cap counter-
clockwise and removing the Air Filter.
Reassemble
in reverse
order
45