Preparation; Application - Vapormed VOLCANO MEDIC Instructions For Use Manual

Table of Contents

Advertisement

7. THE USE OF HEMP FLOWERS

7.2. Preparation

Make sure to use the Mouth-
piece only with attached Lip
Piece before every operation
(see Chapter 9.3. "Disposable MED-
IC VALVE Lip Piece for Hygienic Use",
page 39)
Turn the Filling Chamber Cap counter-
clockwise in order to remove it from
the Filling Chamber.
Remove Filling Chamber Cap
Grind the hemp flowers with the en-
closed herb mill. Place a hazelnut-
sized amount in the Herb Mill and
rotate 4-5 times.
Hemp flowers are ground
most easily when they are
cooled or frozen. The fridge or
freezer are suited for storing hemp
flowers.
Depending on the amount of active
ingredients and the recommended
dosage by a physician, fill the Filling
Chamber with ground hemp flowers.
For consistent and reproduci-
ble dosage with e. g. 100 mg,
use each Filling Chamber
portion only once to fill a Valve Bal-
loon.
Other conditions for a reproducible
dosage are: using standardized can-
nabis, consistent amount, consistent
temperature, consistent Valve Bal-
loon Size (standard size 1.25 l) and
same breathing technique when in-
haling.

7.3. Application

Fill Filling Chamber
Next screw on the Filling Chamber
Cap again.
Screw on Filling Chamber Cap
30
7. THE USE OF HEMP FLOWERS
Take care that the Screens
are not obstructed with plant
material. To guarantee opti-
mal results, we recommend cleaning
the Screens after each vaporization
cycle with the Cleaning Brush sup-
plied with the device.
Suitable for vaporizing small
amounts of hemp flowers, is
the Filling Chamber Reducer,
which is available as an accessory,
along with the Dosing Capsules.
The Filling Chamber Reducer is in-
serted instead of the Filling Chamber
Cylinder (see Chapter 9.5. „Disas-
sembling and Assembling the Filling
Chamber", page 42).
The Filling Chamber Reducer can be
filled with ground hemp flowers direct-
ly. Dosing Capsules containing hemp
flowers can be inserted into the Filling
Chamber Reducer as well.
Fill the Filling Chamber Reducer with
ground hemp flowers directly or with a filled
Dosing Capsule
Press the "HEAT" button to switch on
the heater. When the actual tempera-
ture is equal to the set temperature
put the Filling Chamber with Cap on
the Hot Air Generator and allow it to
lock in place. Put the Valve Balloon
on the Filling Chamber and allow it to
lock in place. This will open the Valve.
Before putting the Valve Balloon on
the Filling Chamber, stretch it tight so
that the Valve Balloon is placed verti-
cally on the Hot Air Generator when it
is filled. Press the "AIR" button. The
Valve Balloon will now fill with the can-
nabis aerosol.
Stretch Valve
Balloon tight
31