A.1 Module 01
Table A.1 defines the Module 01 data available from a Navigator
600 Series analyzer via Profibus cyclic transfer:
Byte
1 to 4
5
6 to 9
10
11 to 14
15
16 to 19
20
21 to 24
25
26 to 29
30
Table A.1 Module 01 Data
6
Type
Description
Float
Stream 1 value
Char
Stream 1 status
Float
Stream 2 value
Char
Stream 2 status
Float
Stream 3 value
Char
Stream 3 status
Float
Stream 4 value
Char
Stream 4 status
Float
Stream 5 value
Char
Stream 5 status
Float
Stream 6 value
Char
Stream 6 status
A.2 Status Byte Definition
Table A.2 defines the meaning of the contents of a Navigator
600 Series analyzer's status byte.
Byte (Hex)
1C
OOS (Out of Service)
80
Good
89
Low alarm limit advisory alarm
8A
High alarm limit advisory alarm
8D
Low-low alarm limit critical alarm
8E
High-high alarm limit critical alarm
Table A.2 Status Byte Definition