Appendix A – GSD File Module
Note.
Float = Floating point number – requires 4 bytes
Char = Character – requires 1 byte
Int
= Integer – requires 2 bytes
A.1 Module 01
Table A.1 defines the Module 01 data available from an Aztec
600 analyzer via PROFIBUS cyclic transfer:
Byte
1 to 4
5
6 to 9
10
11 to 14
15
16 to 19
20
21 to 24
25
26 to 29
30
Table A.1 Module 01 Data
6
Type
Description
Float
Stream 1 value
Char
Stream 1 status
Float
Stream 2 value
Char
Stream 2 status
Float
Stream 3 value
Char
Stream 3 status
Float
Stream 4 value
Char
Stream 4 status
Float
Stream 5 value
Char
Stream 5 status
Float
Stream 6 value
Char
Stream 6 status
A.2 Status Byte Definition
Table A.2 defines the meaning of the contents of an Aztec 600
analyzer's status byte.
Byte (Hex)
1C
OOS (Out of service)
80
Good
89
Low alarm limit advisory alarm
8A
High alarm limit advisory alarm
8D
Low-low alarm limit critical alarm
8E
High-high alarm limit critical alarm
Table A.2 Status Byte Definition