STEP 2. Calibration

To turn the error correction ON, set the calibration type to the full 2-port calibration and
measure the calibration data.
For details about calibration, see Chapter "Calibration" in the E5070B/E5071B User's
Guide.
Step 1. Select the calibration kit suitable for the measurement cable. In this measurement example,
Calibration Kit 85032F is selected.
[Cal] - Cal Kit - 85032F
Step 2. Set the calibration type to the full 2-port calibration using the test port 1 and 2.
[Cal] - Calibrate - 2-Port Cal - Select Ports - 1-2
Step 3. Connect the OPEN standard (included in the calibration kit) to the other end of the
measurement cable that is connected to the test port 1 as shown in Figure 3-13, and
measure the open calibration data at the test port 1. After measuring the open calibration
data, symbol √ is displayed to the left of the
[Cal] - Calibrate - 2-Port Cal - Reflection - Port1 Open
In the same way, measure the calibration data for the SHORT/LOAD standards at the test
port 1.
Figure 3-13
Connecting the OPEN/SHORT/LOAD standards
Step 4. In the same way as described above, measure the calibration data for the
OPEN/SHORT/LOAD standards at the test port 2.
Chapter 3
Measurement Example of a Bandpass Filter
button.