1. To remove the probe tip, hold the probe by
the probe body and twist the probe ring
slightly counter-clockwise. This will loosen the
probe tip.
Caution
ture may dissolve any residual cerumen and
thus contaminate the sensitive parts in the
body of the probe.
2. Take out the probe tip.
3. Check to see if the sound channels of the probe
tip are blocked. You can fit a new probe tip on
the probe, or use the supplied probe tip
flossing thread to clean the sound channels.
Caution
the probe body, as this may cause damage to
the probe.
4. Fit the probe tip on the probe, and twist the
probe ring clockwise to lock the probe tip in
place on the probe.
13.2.4
The test cavities
If a test cavity becomes contaminated, do not use it. Dispose of it and replace it with a new one.
13.2.5
Cleaning the printer
Check the paper compartment regularly to make sure that it is free from paper dust and that the thermal paper head is
clean.
Prerequisites
Before cleaning, switch off MADSEN Zodiac and disconnect it from any external power source.