Cleaning the shoulder strap
Cleaning the shoulder strap ► 36
See
Cleaning the probe tip
Cleaning the probe and probe tip ► 36
See
Disposal
There are no special requirements for the disposal of disposable articles such as eartips and probe tip cleaning floss, i.e.
they can be discarded according to local regulations.
13.2.1
Recommended cleaning agents
Caution
For cleaning the device, we recommend that you only use non-alcohol-based disinfectant wipes (e.g. Audio wipe) or a
cloth dampened lightly with a recommended cleaning agent to ensure proper infection control and maximum lifetime of
the device.
The following chemical solutions are recommended:
Cabinet surface and probes
Non-alcohol-based disinfectant wipes (e.g. Audio wipe)
Ammonium compounds (e.g. dimethyl benzyl ammonium chloride), in concentrations no stronger than 0.1 %.
Aldehyde solutions (e.g. glutaraldehyde),
Oxidizing agents (e.g. Hydrogen peroxide in concentrations no stronger than 3%)
Ortho-phthalaldehyde in concentrations no stronger than 0.6 %.
Caution
13.2.2
Cleaning the shoulder strap
Use a soft, slightly damp cloth with a small amount of cleaning agent to clean the shoulder strap and control pad.
You can remove the shoulder strap from the control pad, If you need to clean the shoulder strap more thoroughly.
13.2.3
Cleaning the probe and probe tip
Although the probes are designed to be easily cleaned, care should be taken to make sure that they last a long time.
Note
Check the sound channels in the probe tip every time you have used the probe.Even small amounts of cerumen
or vernix can block the sound channels. Clean the sound channels if needed.
36
.
Use only the cleaning agents prescribed for cleaning the device.
If plastic parts are soaked in a cleaning agent they will deteriorate.
.