Madsen Zodiac User Manual page 37

Diagnostic & clinical stand-alone auditory impedance tester
Hide thumbs Also See for Zodiac:
Table of Contents

Advertisement

Note
Accurate testing is only guaranteed if you use the eartips approved specifically for MADSEN Zodiac by Oto-
metrics.
Ear canal debris blocking the probe tubes can lead to abnormally large ear canal volume readings, leak messages, and other
odd results. Check the channels of the probe tip every time you use the probe. Even small amounts of cerumen or vernix
can block the probe channels.
Warning
may also be necessary.
Cleaning the probe
Wipe the probe with a disinfectant wipe, such as Audio-wipes, between patients or replace it with a spare one.
Wipe the cable with a disinfectant wipe, such as Audio-wipes.
Wipe the probe home with a disinfectant wipe, such as Audio-wipes.
Alternatively, use a damp, non-flocculent cloth with a small amount of the recommended cleaning agent.
Cleaning or replacing the probe tip
The system is delivered with replacement probe tips. If needed, you can quickly replace a probe tip and clean or discard
the old probe tip at the end of the day.
If the probe tip is only slightly blocked, use the probe tip flossing thread to clean the probe tip channels.
Note
Always comply with local hygienic standards for disinfection.
A. Probe body
B. Probe ring
C. Probe tip
Otometrics - MADSEN Zodiac
Fit a new probe tip on the probe if you have been testing on an infected ear canal. Cleaning the probe ring
User Guide
37