Madsen Zodiac User Manual page 21

Diagnostic & clinical stand-alone auditory impedance tester
Hide thumbs Also See for Zodiac:
Table of Contents

Advertisement

9.2
Using a test setup
In Zodiac
Every time you create a new session in Zodiac, shortcuts to your favorite test setups will be shown on the screen. Use the
soft-buttons to select the test setup you prefer. These test setup short-cuts are shown until you start the measurements.
Selecting a different test setup during a measurement
You can select a test setup different from the one currently selected.
1. Press the
2. Use the arrow buttons to navigate to the desired test setup.
3. Press the
will be used for testing.
9.3
Fitting the eartip on the probe
The Quick Check probe
We recommend that you use an oversized eartip with the Quick Check probe.
The diagnostic probe
Fit the eartip on the probe.
Fitting the eartip on the probe
1. Firmly push and twist the eartip onto the probe tip, until it rests firmly against the base of the probe tip.
Removing the eartip
To remove the eartip, grasp the stem of the eartip and pull the eartip straight off the probe tip.
9.4
Fitting the probe in the patient's ear
1. Look into the ear canal. It is strongly recommended that you perform an otoscopy to assess the status of the outer ear
before you insert the probe.
2. If the ear canal is blocked, this may affect the result of the test. Clean the ear canal if needed.
Caution
Otometrics - MADSEN Zodiac
Menu
button.
Menu
button to exit the menu. The selected test setup
The probe can be damaged if fluids enter the probe.
User Guide
21