5.3 Replacement of Consumables

5.3.1 Replacement of Fuse

WARNING!:
Only use fuses of the rated capacity to prevent injury to personnel
and overheating of the instrument.
WARNING!:
To avoid electric shock, always turn off the power switch and unplug
the AC power cable from the outlet before replacing a fuse.
V-630 uses 2A time-lag fuses (P/N: 5840-H102A 5pcs/set) when using a 100 V voltage supply
and 1.6 A time-lag fuses (P/N: 5840-H106A 5pcs/set) when using a 200 V voltage supply. V-
650/660/670 use 2.5 A time-lag fuses (P/N: 5840-H103A 5pcs/set) when using a 100 V voltage
supply and 1.6 A time-lag fuses (P/N: 5840-H106A 5pcs/set) when using a 200 V voltage supply.
(V-630)
100 V to 120 V:
200 V to 240 V:
(V-650/660/670)
100 V to 120 V:
200 V to 240 V:
2.0 A time-lag fuse
1.6 A time-lag fuse
2.5 A time-lag fuse
1.6 A time-lag fuse
Fuse
Figure 5.1 V-630 rear panel
Fuse
Figure 5.2 V650/660/670 rear panel