Van Der Stahl MS-350 NP Series Operating Instuructions

Microcomputer-controlled sealer for sterilizable bags medical pouch sealer
Hide thumbs Also See for MS-350 NP Series:

Advertisement

Printed AUG 2010
2nd Edition 2.01E
MICROCOMPUTER-CONTROLLED SEALER FOR STERILIZABLE BAGS
MEDICAL POUCH SEALER
MS-350 NP
Series
Operating Instuructions
www.vanderstahl.com

Advertisement

Table of Contents
loading

Summary of Contents for Van Der Stahl MS-350 NP Series

  • Page 1 Printed AUG 2010 2nd Edition 2.01E MICROCOMPUTER-CONTROLLED SEALER FOR STERILIZABLE BAGS MEDICAL POUCH SEALER MS-350 NP Series Operating Instuructions www.vanderstahl.com...
  • Page 2 MS-350 NP Operating Instructions Ver 2.01E...
  • Page 3 Set the heating time according to the material and thickness of the film. As a result of research to further improve the MS-350 NP series, some details in the operating instructions may differ from your actual machine. For operating problems or product information, please contact your local dealer.
  • Page 4: Table Of Contents

    MS-350 NP Operating Instructions Ver 2.01E Table of contents Specification ………………………………………………………………………………………… 5 Accessories …………………………………………………………………………………………… 5 Replacement parts …………………………………………………………………………………… 6 Proper use of the sealer …………………………………………………………………………… 6 Warning-caution label location and content …………………………………………………… 7 Operating precautions ……………………………………………………………………………… 8 For safety operation…………………………………………………………………………………………………………………………… 8 For smooth operation ………………………………………………………………………………………………………………………...
  • Page 5: Specification

    MS-350 NP Operating Instructions Ver 2.01E 1 Specification Machine Voltage Power Sealing Sealing Weight Code Model Dimension mm Width mm Length mm W x D x H MS-350 NP 1500 525x465x200 2 Accessories Please make sure that all of the accessories described below are included. If not, please contact your dealer.
  • Page 6: Replacement Parts

    Film thickness limits for the MS-350 NP series are described below. The machine cannot seal bags which are thicker than that indicated. The machine may malfunction if you try to exceed the limits of the machine.
  • Page 7: Warning-Caution Label Location And Content

    WARNING electric shock UNPLUG BEFORE PERFORMING MAINTENANCE. CAUTION KEEP HANDS CAUTION AWAY If Warning - Caution labels peel off or are lost, please purchase the proper la- bels and place them in the appropriate location.
  • Page 8: Operating Precautions

    Do not modify the machine Warning Modification of the MS-350 NP series or the use of unspecified parts may not only lead to damage of the machine, but there is also a risk of danger to the operator. Please do not tamper with the machine. Accidents or damage caused by modification by the customer is the sole responsibility of the customer.
  • Page 9 MS-350 NP Operating Instructions Ver 2.01E Do not use parts other than those specified Warning When replacing parts, always use only specified parts sold through Fuji Impulse. Not responsible for malfunctions caused by the use of unspecified parts. If the power switch automatically turns to the OFF Warning position If after 3 seconds heating is not complete after it has been initiated, the...
  • Page 10: For Smooth Operation

    MS-350 NP Operating Instructions Ver 2.01E For smooth operation ● Heating temperature should be set at a minimum Set sealer so that an acceptable seal is completed at the lowest heating temperature. Excessively high heating temperatures will damage the Teflon, glass tape and other consumables, as well as result in heating element breakage.
  • Page 11: Other Safeguards

    ● The Fuji Impulse MS-350 NP series is an impulse sealer Do not become alarmed if the sealing area does not become hot after the power switch is turned ON. The machine has not malfunctioned.
  • Page 12: Major Parts And Function

    MS-350 NP Operating Instructions Ver 2.01E 7 Major parts and function Frame cover Control unit 1. Set the seal condition according to the film being used. 2. Set the auto-cycle interval. 3. Turn ON or OFF the auto cycle interval mode.
  • Page 13 MS-350 NP Operating Instructions Ver 2.01E...
  • Page 14: Proper Use Of The Sealer

    MS-350 NP Operating Instructions Ver 2.01E 8 Proper use of the sealer Use only as described in this manual. There is a risk of danger and malfunctions may result if the machine is used not in accordance with this manual. 8-1 Setting the machine Install the machine on a flat, secure working place.
  • Page 15: Power Switch On

    MS-350 NP Operating Instructions Ver 2.01E 8-4 Power switch ON Turn ON the power switch located on the side of the machine. The digital display on the control panel will light up. (The number initially displayed is the counter.) Push to turn ON 8-5 Select the operation mode The "Auto cycle START/STOP"...
  • Page 16: Setting The Sealing Condition With The Control Unit

    MS-350 NP Operating Instructions Ver 2.01E 8-6 Setting the sealing condition with the control unit DIGITAL DISPLAY AUTO CYCLE START/STOP Counter and setting display ・ The heating temperature, heating time (heating temperature retaining time), cooling temperature and auto-cycle interval are set with the control unit. The setting values will vary depending on the pouch/film material, thickness, etc.
  • Page 17 MS-350 NP Operating Instructions Ver 2.01E Setting the different modes MODE DIGITAL DISPLAY Power ON TEMP. DISPLAY DECREASE VALUE INCREASE VALUE PUSH - FIRST TIME ....HEATING TEMP. DECREASE VALUE INCREASE VALUE PUSH - SECOND TIME ....HEATING (RETAINING) TIME PUSH - THIRD TIME ....
  • Page 18: Adjust The Seal Pressure

    MS-350 NP Operating Instructions Ver 2.01E 8-7 Adjust the seal pressure The pressure adjusting nut is adjusted to the "NORMAL" position at the time of factory shipment. The required sealing pressure varies depending on the type and thickness of the film. Adjust the sealing pressure according to each of the film to be used.
  • Page 19: Finishing Sealing

    MS-350 NP Operating Instructions Ver 2.01E ● ABORT CYCLE button To stop the operation after the machine started working, push the “ABORT CYCLE” button. The lever will return to the initial position. The set value will not be erased even if the “ABORT CYCLE” button is pressed. 8-10 Finishing sealing After the heating and cooling cycle has finished, the pressure lever returns to the initial position, and the...
  • Page 20: Power Switch Off

    MS-350 NP Operating Instructions Ver 2.01E 8-12 Power switch OFF After finishing operation, always turn OFF the power switch. Push to turn OFF As the set value and the seal count are retained by the control unit, there is no need to put the value every time after turning OFF the power switch.
  • Page 21: Safety Devices And Error Messages

    MS-350 NP Operating Instructions Ver 2.01E 9 Safety devices and error messages When pinching the fingers to the sealing ● area When the foreign objects such as fingers are caught in the sealing area, the microswitch located under the pressure lever is not activated and the pressure lever returns to the initial position.
  • Page 22: Replacing The Routine Maintenance Parts

    MS-350 NP Operating Instructions Ver 2.01E 10 Replacing the routine maintenance parts Daily maintenance or replacement of parts at an early stage can prevent the machine from being damaged, and keeps the machine at the optimum condition. Replace parts according to the following directions in order to maintain your machine in optimum operating condition.
  • Page 23: Preparation

    MS-350 NP Operating Instructions Ver 2.01E 10-1 Preparation ●Before replacing parts... When replacing parts, lift up the pressure lever for easier replacement. Pressure adjusting nut How to lift up the pressure lever Pressure adjusting nut Loosen the M5x10 frame cover fixing screws, fastening screw and remove the frame cover.
  • Page 24: Replacing The Teflon Center-Dry Tape

    MS-350 NP Operating Instructions Ver 2.01E 10-3 Replacing the Teflon center- dry tape Essential tools: Scissors, Philips screwdriver Replace when: The center-dry tape breaks, burns, or when the seal becomes messy, etc. The Teflon center-dry tape is sold individually. Non-adhesive area Remove the old center-dry tape, starting from the edge.
  • Page 25: Replacing The Temperature Sensor

    MS-350 NP Operating Instructions Ver 2.01E 10-4 Replacing the temperature sensor Sarcon sheet Essential tools: A Philips screwdriver Pressure lever Replace when: The temperature sensor is damaged. The temperature sensor is also sold individually. The temperature sensor is positioned so that it slides in Glass tape between the heating element and glass tape.
  • Page 26: Replacing The Heating Element

    MS-350 NP Operating Instructions Ver 2.01E 10-5 Replacing the heating element Electrodes that mount the heating element consist of Electrode mounding parts as in the right illustration. bracket Electrode plate Essential tools: A Philips screwdriver Electrode plate spring Replace when: The heating element breaks, Electrode cover Screw D unevenness is generated, or when the...
  • Page 27: Replacing The 38Mm-Width Glass Tape And Sarcon Sheet

    MS-350 NP Operating Instructions Ver 2.01E 10-6 Replacing the 38mm-Width glass tape and Sarcon sheet Essential tools: Scissors, A Philips screwdriver Replace when: The heating element breaks often, the seal becomes messy, etc. The glass tape and Sarcon sheet are either sold 360mm individually or included in the maintenance parts kit.
  • Page 28: Replacing The Silicone Rubber (White)

    MS-350 NP Operating Instructions Ver 2.01E 10-7 Replacing the silicone rubber (white) Essential tools: Alcohol (Ethanol) Replace when: The seal becomes messy, etc. Silicone rubber (white) The silicone rubber is either sold individually or included in the maintenance parts kit. Remove the 38mm-width glass tape adhered on the silicone rubber.
  • Page 29: Replacing The Microswitch

    MS-350 NP Operating Instructions Ver 2.01E 10-9 Replacing the microswitch Essential tools: A Philips screwdriver Replace when: The problems listed on "13 Common problems and solutions" occurs. The microswitch is sold individually. Always unplug the power cord from the Warning outlet when replacing the microswitch.
  • Page 30: Adjusting Each Part

    MS-350 NP Operating Instructions Ver 2.01E 11 Adjusting each part 11-1 Adjusting the seal pressure Adjust the pressure strength indicated on the seal pressure adjusting nut using the pointer on the small lever. NORMAL Loosen the pressure adjusting nut fastening screw.
  • Page 31: Electric Diagram

    MS-350 NP Operating Instructions Ver 2.01E 12 Electric diagram 110V MS-350 WIRING DIAGRAM Gray Gray Microcomputer control unit Yellow Black Black Black White Gray Black Transformer TRIAC Gray Electrode Heating element Temperature sensor Power cord Microswitch Earth black Yellow Terminal strip Gray Black Gray...
  • Page 32: Common Problems And Solutions

    MS-350 NP Operating Instructions Ver 2.01E 13 Common problems and solutions If you are experiencing problems with the machine, please refer to the chart below. Carefully follow the directions in the operating instructions when replacing parts. Please consult your dealer or Fuji Impulse if after referring to the chart, the problem cannot be resolved. Please refer any questions regarding replacement of parts not listed in the operating instructions or adjustment of such parts to your dealer or Fuji Impulse.
  • Page 33 MS-350 NP Operating Instructions Ver 2.01E Problems Check Solutions Is the heating element broken? Replace the heating element. Polish the metal area where the Bad connection between the heating electrode and heating element element and electrode. contact using the sand pater. Heating lamps turns on, but there Is the black (blue) wire from the is no heat and sealing cannot be...
  • Page 34: Error Message And Solutions

    MS-350 NP Operating Instructions Ver 2.01E 14 Error message and solutions Error number Check Solutions ・This error is displayed when the microswitch is damaged. Replace the → Is the microswitch damaged? microswith or please consult your local dealer. ・If the lever continues to be in the →...
  • Page 35: Password Function

    MS-350 NP Operating Instructions Ver 2.01E 15 Password Function Password Function Password function is installed for the control of sealing conditions - “Heat temp.” “Heat temp-keeping time” “Cool temp.” and “Auto cycle interval”. It protects the sealing condition settings from being changed by the other person excepting main user. A.
  • Page 36 MS-350 NP Operating Instructions Ver 2.01E Set the sealing condition - When the input password is correct, HEAT TEMP. SETTING lamp turns on and display shows the present heating temperature. Setting mode is alive. - Select the new temperature by UP / DOWN button and go to the next by SET button. - After the all condition settings, seal test can be done.
  • Page 37: Exploded View And Parts List

    MS-350 NP Operating Instructions Ver 2.01E Exploded View Spare Parts list When ordering spare parts, please consult the exploded view diagram/spare parts list on the following pages and provide both the part name and code number to your local dealer.
  • Page 38 E Type MS-350 NP Exploded View MS-350 NP �� � � � � �� � �� �� � �� ��� �� �� � �� �� �� � �� �� �� �� �� � �� �� �� �� � �� �� �...
  • Page 39 E Type MS-350 NP Parts List MS-350 NP ● Please indicate the part name and number in be below list when placing your order. ● Please specify the model name of your sealer and parts name when no code no. is listed. Sales Sales Parts...
  • Page 40 Y6602 AUG 2010 2nd Edition MS-350 NP Operating Instructions Ver 2.01E...

Table of Contents