PSS® 7000 SERIES c/w SENTINEL 7000 EMU
Check that the face mask port and the connector O-ring of the LDR are clean and
undamaged.
Align and push the LDR into face mask port until it latches in position.
Check the attachment by attempting to pull the LDR away from the face mask.
Normal Operation
All safety information, including the preliminary safety details, must be read and
!
understood before commencing any user task. The procedures detailed in this
UI may only be carried out by trained and competent SCBA users.
Before use checks
Carry out a visual inspection of the complete SCBA including any ancillary equipment that
will be used during the operation. The tasks detailed in the preparation for use and the pre-
operational checks sections must also be completed satisfactorily before release of the SCBA
for operational use. Components that are worn or damaged must be repaired or replaced
before proceeding. Do not use the equipment if any of the following is found:
The SCBA fails to meet any of the test specifications.
Any casing is cracked or broken.
The LCD display of the user interface cannot be read, or one or more of the display
segments is inoperative.
Audible and/or visual alarms are not functioning.
Any fault icon shown.
Sentinel 01: Edition 1: July 2008 - Subject to Modification
16