PSS® 7000 SERIES c/w SENTINEL 7000 EMU
i
i
i
i
psi
approximately one second intervals
to confirm that the Sentinel 7000
has passed the self check and is in
the active mode.
If a cross icon (Fig 12) with
i
i
a fault code is displayed, the
Sentinel 7000 has failed the
self check. Contact Dräger
for repair.
Press and hold the RH and LH
buttons of the user interface until
the display clears, then immediately
release the buttons.
High Pressure Leak Test
and EOSTI Warning Test
The following is not intended as
a test of HUD functionality. If the
HUD fails to operate as described,
refer to the HUD UI (instruction no.
3356234).
Place the face mask next to the
SCBA (within three feet of the
pressure module).
Press the reset button (1, Fig 13)
of the LDR to switch off the positive
pressure.
Sentinel 01: Edition 1: July 2008 - Subject to Modification