Continuity Testing (Resistance Tests) - KYORITSU 3132A Instruction Manual

Analogue insulation-continuity tester
Table of Contents

Advertisement

required test voltage when providing a steady state current of 1mA.
The minimum allowable resistance level is 0.25MΩ for the 250V test,
0.5MΩ for the 500V test and 1MΩ for the 1000V test.
● Principle of Insulation Resistance Measurement
Resistance value can be obtained by applying a
certain high voltage to the resistance
(insulation resistance) and measuring the
flowing current.
Resistance Value = Voltage / Current

6-3 Continuity Testing (Resistance Tests)

● Always test the circuit or equipment to ensure it is surely de-energized
before measurement according to the instruction of 6-1.
● To avoid electrical shock, measurements must be performed on de-
energized circuits only.
● Never make measurement with the battery compartment cover
removed.
RX
= V / I
DANGER
― 13 ―