1.
Remove the cable tie A.
2.
Remove the contact disc cables B from the Mover PCB, using
a terminal screwdriver. Note brown cable top and blue cable
bottom.
Feed the cables through the chassis plate.
3.
Carefully separate and remove the complete contact disc C
from the suspension base D by releasing the self-adhesive tape.
4.
Remove, clean off the remains of the self adhesive tape.
1586747-A
5.
Fit the new self adhesive tape E and the new contact disc C.
6.
Feed the cables B through the chassis plate and connect them
to the Mover PCB.
7.
Re-assemble the hoist and reset the emergency stop.