Replacing The Fuse; Adjusting The Solenoid Until The Hoist Traverses - Invacare Robin Service Manual

Ceiling hoist
Hide thumbs Also See for Robin:
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 1

3.8 Replacing the fuse

CAUTION!
Static electricity can damage the PCBs inside the hoist.
Before working in the inside of the hoist:
– Pull the red emergency strap to turn off the hoist.
– Remove the hoist from the track system. (Refer to 3.2
Dismounting the ceiling hoist, page 9 )
– Remove the cover from the hoist. (Refer to 3.4
Removing the hoist cover, page 12)
– An anti-static mat must be used when working inside
the ceiling hoist.
1.
Remove the hand control. Refer to chapter 3.3 Replacing the
hand control, page 12
2.
Remove the battery. Refer to steps 1 to 3 in chapter 3.6
Replacing the battery, page 13.
Gently pull the fuse A straight out away from the holder B.
3.
4.
Insert new fuse.
Make sure only a fuse with correct rating 10A 250V
slow blow is used.
5.
Re-insert battery, re-assemble the hoist, insert the hand control
and reset the emergency stop.
1586747-A
3.9 Adjusting the solenoid until the hoist
traverses
Robin Mover™ only.
4 mm Allen key; 7 mm spanner
1.
Pull the red emergency strap.
2.
With the ceiling hoist remaining on the track, remove both
plastic shells:
a.
Remove the screws securing the plastic shells from the rail
motor.
b.
Pull the plastic shells away from each other to remove them
from the rail motor.
Service