Three phase apparent power
Equivalent three phase
power factor
8.6.3
HARMONIC FORMULAE
Total Harmonic Distortion
Factor of Voltage
Total Harmonic Distortion
Factor of Current
where
Cos∅ fnd = Cosinus of the phase angle between V1 and A1
8.7
PRINTER SPECIFICATIONS
• Number of columns:
42.
• Characters:
5x7 matrix
• Print speed:
1 line per second.
• Paper:
2
55g/m
pure cellulose smooth.
• Paper width:
57 mm.
• Paper length:
16 m.
• Print functions:
Manual (PRINT key on instrument's front panel prints out all measurements taken).
Automatic (printouts generated automatically at time intervals of 1 to 99 minutes as programmed on calendar-
clock). Congruity of printed measurements ensured by reference to start of printout time.
8.8
CLAMP METER SPECIFICATIONS
• Measurement range:
from 0,1 A to 1200 A.
• Frequency range:
from 30 Hz to 10 kHz.
• Ratio:
1000A/1Vrms.
• Accuracy:
200... 1000A ≤ 0,5%
200A ≤ 0,75% ≤ 0,75° phase angle
50...
50A ≤ 1,5%
10...
10A ≤ 3% + 0,1 mV
0,1...
V
, A
= V, I
harmonic of 50/60 Hz fundamental
k
k
k = 2, 3,..., 24th harmonic computed by means of a DFT
≤ 0,5° phase angle
≤ 1,5°
phase angle
VA
=
W
∑
CosΦ
= W
∑
∑
THDF V =
THDF A =
40
+
2
2
VAr
∑
∑
/ VA
∑
24
24
∑
2
V
∑
V
or
k
k
2
2
V
V
1
rms
24
24
2
2
∑
∑
A
A
k
or
k
2
2
A
A
rms
1
2
Need help?
Do you have a question about the Microvip 3 plus and is the answer not in the manual?