Power Requirements - Thermo Scientific TriPlus RSH Preinstallation Manual

Robotic sample handling
Hide thumbs Also See for TriPlus RSH:
Table of Contents

Advertisement

1
TriPlus RSH Site Preparation

Power Requirements

Table 5.
Overall Dimensions of the GC System + TriPlus RSH with Extended X-axis Configuration (Sheet 2 of 2)
Configuration
FOCUS GC + ISQ MS + TRACE GC Ultra + TriPlus RSH
FOCUS GC + ISQ Series MS + FOCUS + TriPlus RSH
TRACE GC Ultra + DSQ II MS + TRACE GC Ultra + TriPlus RSH 89
FOCUS GC + DSQ II MS + TRACE GC Ultra + TriPlus RSH
FOCUS GC + DSQ II MS + FOCUS + TriPlus RSH
TRACE GC Ultra + ITQ MS + TRACE GC Ultra + TriPlus RSH
FOCUS GC + ITQ MS + TRACE GC Ultra + TriPlus RSH
FOCUS GC + ITQ MS + FOCUS + TriPlus RSH
TRACE GC Ultra + TRACE GC Ultra +
FOCUS GC + TRACE GC Ultra + TSQ MS + TriPlus RSH
FOCUS GC + FOCUS + TSQ MS + TriPlus RSH
Power Requirements
Power Line
Power Module
8
TriPlus RSH Preinstallation Requirements Guide
TSQ MS +
This paragraph gives details on the power requirements.
CAUTION The power line and the connections between the instruments must maintain
good electrical grounding. Poor grounding represents a danger for the operator and might
seriously affect the instrument performance.
Do not connect the TriPlus RSH to lines feeding devices of a heavy-duty nature, such as
motors, UV lamps, refrigerators, air compressors and other devices that can generate
disturbances. Pay attention not to leave any cable connecting the sampling unit and the
chromatographic system or the power cord close to the GC hot air vents.
Occasionally, unacceptable quality in line power sources might adversely affect the
operation of a GC system. It is the user's responsibility to correct line voltage problems.
Specifying power conditioning equipment is a complex task that is best handled by a
company or consultant specializing in that field. Contact your Thermo Fisher Scientific
Field Service Engineer (FSE) for assistance in locating a power consultant.
The TriPlus RSH must be electrically supplied by using a dedicated power supply module. A
second Power module is required in case a Stack modules (Peltier) is installed. Do not use the
same Power module as used for the TriPlus robot.
Depth
89
35
cm
in.
89
35
35
89
35
89
35
89
35
89
35
89
35
TriPlus RSH
89
35
89
35
89
35
Width
Height
135
53
127
50
cm
in.
cm
in.
135
53
127
50
162
64
127
50
137
54
127
50
135
53
127
50
162
64
127
50
140
55
127
50
135
53
127
50
210
82.6
127
50
170
67
127
50
210
82.6
127
50
Mass
150
331
kg
lbs
132
292
168
371
150
331
132
292
168
371
150
331
150
331
241
532
223
492
205
452