Drx Detector Led Display Matrix - Carestream DRX-1 User Manual

Drx family
Table of Contents

Advertisement

Detector Operation

DRX Detector LED Display Matrix

LED Pattern
1 Green Flash
2 Green Flashes
3 Green Flashes
4 Green Flashes
5 Green Flashes
1 Blue Flash
2 Blue Flashes
3 Blue Flashes
4 Blue Flashes
5 Blue Flashes
2–14
The meaning of the LED Pattern.
The detector is in Standby, and no
study is active. The Console may
not be connected, or the power is
low.
The System is on, and in Ready
status.
The System is on and enabled, but
not ready.
The System is on and enabled.
These flashes appear while press-
ing the prep switch.
A power fault has occurred. The
detector is unusable.
The detector is currently con-
nected to the System Console.
The detector is not currently con-
nected to the System Console.
The detector has rebooted and
has not yet connected to the Sys-
tem Console.
The detector is performing a firm-
ware upload.
The detector is attempting to con-
nect to the wireless Access Point
(AP).
The action the user should take.
No action is required. The detector
is not being used.
No action is required. The LED
indicates that the detector has
been selected.
This should not display in normal
operation. If this pattern is visible,
call Service.
No action is required.
Remove and re-insert the battery
to clear the problem. If the prob-
lem persists, call Service.
No action is required.
If this pattern occurs while detec-
tor is in use, call Service.
If this pattern occurs when the
detector should be in use, call Ser-
vice.
Do not interact with the detector.
No action is required.
AA5348 | 2012-12-20