Detector Operation

Shipping and Handling Labels

The following graphic is applied to the shipping package, and describes the conditions
that should be met while the package is in transit and storage. It is shown for reference
only. The actual shipping label may vary slightly.
Figure 13: Shipping label
Figure 14: Detector Weight Limit Label
Caution
Since the detector is not a patient support device, it must be placed on a suitable surface, such as
a table or floor, before applying patient weight to it. The weight label indicates acceptable limits
of use that will not damage the detector. To prolong the life of the detector, and minimize
potential internal detector damage, observe the following weight restrictions:
The maximum concentrated weight over a small area of the detector surface (50 mm (2 in.))
in diameter must not exceed 114 kg (250 lb).
The maximum distributed weight applied uniformly over the entire detector surface is 170 kg
(375 lb).