Recommended EMC Environment

Guidance and Manufacturer's Declaration—Electromagnetic Emissions
The detector is intended for use in the electromagnetic environment specified below. The customer or
the user of the detector should assure that it is used in such an environment.
Emissions Test
RF Emissions
CISPR 11
RF Emissions
CISPR 11
Harmonics Emissions
IEC 61000–3–2
Voltage
Fluctuations/
Flicker Emissions
IEC 61000–3–3
Electromagnetic Immunity for Equipment and Detectors Fully Compliant with IEC 60601–1–2:2007
The detector is intended for use in the electromagnetic environment specified below. The customer or
the user of the detector should assure that it is used in such an environment.
Immunity Test
Electrostatic Discharge
(ESD)
IEC 61000–4–2
1–6
Compliance
Electromagnetic Environment—Guidance
Group 1
The detector uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely
to cause any interference in nearby electronic equipment.
Class A
The detector is suitable for use in all establishments other
than domestic and those directly connected to the public
low-voltage power supply network that supplies buildings
used for domestic purposes.
Class A
Complies
IEC 60601
Compliance
Test Level
Level
± 6 kV contact
± 6 kV contact
± 8 kV air
± 8 kV air
Electromagnetic
Environment—Guidance
Floors should be wood, concrete, or
ceramic tile. If floors are covered with
synthetic material, the relative
humidity should be at least 30 %.
AA5348 | 2012-12-20