Read Transducer A/D Counts - Pressure Systems 9022 User Manual

Netscanner system intelligent pressure scanner
Table of Contents

Advertisement

Pressure Systems, Inc.
READ TRANSDUCER A/D COUNTS (Command 'a')
Purpose:
Returns the most recently acquired raw pressure data for the specified channels in
averaged signed A/D counts (in the range -32768 to +32767). This simple data
bypasses any usage of the transducer's factory-calculated coefficients or the final
calibration process's adjustment coefficients (offset and gain). Each datum returned
in the response will be in the specified high-precision data format, but representing
A/D counts as a signed integer average. (The formula for converting A/D counts to
volts is: Volts = A/D Counts * 5/32768) This command is intended for advanced
users only and is not required for normal operation.
Command
"appppf"
'a' is the command letter
'pppp' is the position field
'f' is the format field
Response
" dddd.. dddd"
' dddd' are the data fields, each with leading space (except f = 7 or 8).
Description: The 4-character hex position field (pppp) specifies a 16-bit binary bit-map, with
each bit (set to 1) specifying a particular channel number (16-1, left-to-right). Only
channels 12-1 are allowed for Models 9021 and 9022.
The 1-character format field (f) specifies the format of each data field (dddd) that
will be returned in the requested response. The first datum returned in the response
will be for the highest channel number requested. Each datum will be preceded by
a space character. Some formats may not be applicable to the specific type of data
being requested. Valid formats are shown in the following table:
f
converts each internal response datum value from..
0
single binary float to
1
single binary float to
2
double binary float
5
single binary float to
7
single binary float to
8
single binary float to
NetScanner™ System (9016, 9021, & 9022) User's Manual
7-10-digit signed decimal " [-xxx]x.xxxxxx"
8-digit hex integer " xxxxxxxx"
to
16-digit hex integer " xxxxxxxxxxxxxxxx"
long integer (EU*1000) then to 8-digit hex integer
single binary float (big endian: msb first)
single binary float (little endian: lsb first)
3 - 28
max.char.
13
9
17
9
4
4
www.PressureSystems.com

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

90169021

Table of Contents