Trademarks The Bluetooth® word mark and logos are registered trademarks owned by Bluetooth SIG, Inc. and any use of such marks by i-SENS, Inc. is under license. All other trademarks and trade names are those of their respective owners. Note: The SmartLog mobile app may not be compatible with all Smartphones.
Page 3
CareSens N Test Strips. should be used only for self-testing outside the body (in vitro diagnostic use only). CareSens N Premier Blood Glucose Monitoring System should not be used for the diagnosis of level over 65% or below 15%) may produce inaccurate diabetes or for testing newborns.
Specifications CareSens N Premier Blood Glucose Monitoring System Product specifications CareSens N Premier BGM System includes the following items: * CareSens N Premier Blood Glucose Meter Measurement range 20-600 mg/dL (1.1-33.3 mmol/L) * Owner's Booklet Sample size Minimum 0.5 μL...
Inserting or Replacing the Batteries Caring for Your System The CareSens N Premier meter uses two 3.0 V lithium batteries. Use a soft cloth or tissue to wipe the meter exterior. If necessary, dip the soft cloth or tissue in a small amount of alcohol.
CareSens N Blood Glucose Test Strip Warning! The CareSens N Premier Blood Glucose Monitoring System measures blood glucose quickly and accurately. It automatically absorbs the small blood sample applied to the narrow edge of capillary whole blood samples. the strip.
Page 7
CareSens N Premier Blood Glucose Meter CareSens N Premier Blood Glucose Meter Display Test results : test results displaying panel Data Port Display Memory recall mode : appears when Used to transfer data Shows results, test results stored in the memory are...
Before pairing your meter and Bluetooth® Smart feature is off. When smartphone, download and install the SmartLog™ mobile you need to turn off/on the Bluetooth® Smart feature, press the app on your smartphone. button when OFF/On blinks on the screen. 16 www.i-sens.com www.i-sens.com 17...
Page 9
Note: PIN number. Some smartphones, especially those that are not tested or approved by i-SENS, may be incompatible with your meter. Visit www.i-sens.com/smartlog for more information about supported smartphones. You can also scan the QR code on the back cover of this user manual.
Press the ◀ or ▶ button until the correct displays the correct date. Press the button to minut appears. After setting the minute, confirm the date and to go to the next step. press the button to go to the next step. 20 www.i-sens.com www.i-sens.com 21...
If the expiration date is set to October of 2017, the meter Note: will display EP at the start of symbol is displayed only when the sound is set to November, 2017. OFF. 22 www.i-sens.com www.i-sens.com 23...
After corner. the following. setting, press and hold the button for 3 seconds to turn off the meter. Caution: Ask your healthcare professional to help you decide what your hypoglycemia level is before setting your level. 24 www.i-sens.com www.i-sens.com 25...
Press pressing the ▶ button for temperature between 8-30˚C (46-86°F). 3 seconds. To undo the hold ▶ control solution flag, press the ▶ button for 3 seconds again. 26 www.i-sens.com www.i-sens.com 27...
Page 14
The range printed on the test strip vial is for the CareSens Control Solution only. It has nothing to do with your blood glucose level. Note: The CareSens Control Solution can be purchased separately. Please contact your authorised i-SENS sales representative. 28 www.i-sens.com www.i-sens.com 29...
If results continue to fall outside the range printed on the test Note: strip vial, the CareSens N Test Strip and CareSens N Premier Repeated puncturing at the same sample site may cause pain meter may not be working properly. Do not use your system or skin calluses (thick hard skin).
The skin depth to get blood samples will vary for various with the load confirmation window and people at different sample sites. The lancing device's release button as shown. adjustable tip allows the best depth of skin penetration to get an adequate sample size. 32 www.i-sens.com www.i-sens.com 33...
Wait a few and start from Step 2. seconds for a blood drop to form. A minimum volume of 0.5 microliter is needed to fill the confirmation window. (Actual size of 0.5 μL: ) 34 www.i-sens.com www.i-sens.com 35...
Page 18
Control solution flag The SmartLog mobile app must be launched. The meter will transmit data in the following cases, When the strip is ejected after measuring, When the meter is turned on (only when untransmitted data exists). 36 www.i-sens.com www.i-sens.com 37...
Then press the release button while continuing to apply Always dispose of lancets properly. pressure. Keep holding the lancing device against your skin until sufficient (at least 0.5 μL, actual size: ) blood is drawn. Carefully lift the lancing device away from your skin. 38 www.i-sens.com www.i-sens.com 39...
Page 20
Try puncturing again in a smoother area. 40 www.i-sens.com www.i-sens.com 41...
Diabetes -2015. ” Diabetes Care. January 2015; 38(1):S10, S81. please contact your healthcare professional immediately. Transferring Test Results Using Cable Test results stored in CareSens N Premier meter can Note: be transferred from the meter to a computer using SmartLog software and cable. The meter screen displays such messages are displayed even though you do not have "Pc"...
Meter Memory The CareSens N Premier meter can save up to 1,000 glucose test Step 3 Viewing Pre-meal Averages results with time and date. If the memory is full, the oldest test button Repeatedly press the ◀ result will be deleted and the latest test result will be stored.
Page 23
After checking the stored test results hold the button to turn off the meter. Note: The control solution test results saved with symbol will be displayed with symbol when you review the stored test results. 46 www.i-sens.com www.i-sens.com 47...
Setting the Alarm Function Step 2 Turning the PP2 alarm OFF Four types of alarms can be set in the CareSens N Premier To turn off the PP2 alarm, press Meter: one post-meal alarm (PP2 alarm) and three time set and hold the ◀...
Pres the ◀ button to adjust the time of Press the button for 3 seconds to finish and turn the meter ‘alarm1’ . off. A number representing the hour will blink on the screen. Press the ▶ button to set the hour. 50 www.i-sens.com www.i-sens.com 51...
The blood sample has abnormally high viscosity or insufficient volume. > Repeat the test after inserting a new test strip. Note: If the error messages persist, contact your authorised i-SENS sales representative. 52 www.i-sens.com www.i-sens.com 53...
Monitoring System has been evaluated in laboratory and in clinical tests. inserted with the contact bars facing Accuracy: The accuracy of the CareSens N Premier BGM System up. Check if the strip has been (Model GM01AAB) was assessed by comparing blood glucose...
Page 28
Interval 2 Interferent Precision: The precision studies were performed in a laboratory 50~100 mg/dL 250~350 mg/dL (2.8~5.5 mmol/L) (13.9~19.4 mmol/L) using CareSens N Premier BGM Systems. Acetaminophen -2.3 mg/dL (-0.1 mmo/L) -3.3% Within Run Precision *Bloodav 44 mg/dL (2.4 mmol/L) SD = 2.1 mg/dL (0.1 mmo/L)
Warranty Information Manufacturer’s Warranty Difference Averages i-SENS, Inc. warrants that the CareSens N Premier Meter shall Interval 1 Interval 2 Interferent be free of defects in material and workmanship in normal 50~100 mg/dL 250~350 mg/dL use for a period of five (5) years. The meter must have been (2.8~5.5 mmol/L)