Measurement - Lutron Electronics CL-2006 Operation Manual

Chlorine meter
Table of Contents

Advertisement

3) Insert the " Testing bottle " to the bottom of
" Container " ( 3-2, Fig. 1 ) completely.
4) Before the measurement, it should envelop in the
" Cover " ( 3-1, Fig. 1 ) completely.
Remark :

4-2 Measurement

Before the measurement, should select the
measurement model to " Free Chlorine " or " Total
Chlorine " first, the setting procedures refer to
chapter 4-3, page 8.
ZERO setting for the liquid
1) Power ON the meter by pressing the " Power Button "
( 3-6, Fig. 1 ) once.
The Display ( 3-3, Fig. 1 ) will show the text " Free (
TOTAL ) " approx. 1 second. then show " CAL 0 ", now
the meter for the ZERO setting procedures.
2) Fill the measurement liquid into the " Testing bottle "
( 3-11, 3-12, Fig. 1 ) .
Attention :
It should fill the
measurement
liquid capacity
until its level
up to the " level
marker " of the
" Testing bottle "
Before the measurement, it should
keep the outside of Testing Bottle
under the dry condition and without
existing any dust.
6
Model 0601
10 mL
Fig. 3