5. CALIBRATION PROCEDURE
1) The meter can be calibrated under following calibration
points :
Free chlorine
* Zero
* 1.00 ppm
Total chlorine
* Zero
* 1.00 ppm
2) The meter ship along with
* Zero Chlorine standard solution ( CL-01 ) X 1 PC,
* 1.0 ppm Free Chlorine standard solution ( CF-01 ) X 1 PC.
* 1.0 ppm Total Chlorine standard solution ( CT-01) X 1 PC.
as the standard accessories.
3) The complete calibration should be executed by
following two solution :
* Zero Chlorine standard solution
* 1.0 ppm standard solution ( Free or Total ).
or
10