Performance Verification
4-14
8. Touch the SDI soft key to set the video measurement set to Digital mode.
9. In the Measurement mode display, touch the SDI Format application.
10. Verify that no error messages appears on the SDI Format Monitor display.
11. Press the Picture button to set the video measurement set to SDI Picture
application.
12. Verify that the correct color bars and overlay logo are displayed on the SDI
Picture display.
13. Press the Measure button on the video measurement set to open the Measure
mode display.
14. In the Measure mode display, touch the SDI Audio Format application.
15. Verify that no CRC errors appear on the SDI Audio Format Analyzer display.
16. Move the BNC cable from the SDI 1 connector to the SDI 2 connector on the
sync pulse generator.
17. Press the Measure button on the video measurement set to open the Measure
mode display.
18. Repeat steps 9 through 15.
SPG600 Option 03 Only. Perform steps 19 through 24 if Option 03 is installed in
your SPG600.
19. Move the BNC cable from the SDI 2 connector to the SDI 3 connector on the
sync pulse generator.
20. Press the Measure button on the video measurement set to open the Measure
mode display.
21. Repeat steps 9 through 15.
22. Move the BNC cable from the SDI 3 connector to the SDI 4 connector on the
sync pulse generator.
23. Press the Measure button on the video measurement set to open the Measure
mode display.
24. Repeat steps 9 through 15.
SPG600 & SPG300 Sync Pulse Generators Service Manual