Tektronix SPG600 Service Manual page 83

Sync pulse generators
Hide thumbs Also See for SPG600:
Table of Contents

Advertisement

SPG600 & SPG300 Sync Pulse Generators Service Manual
24. Move the BNC cable connection from the CH 4 (or CH 8) connector to the CH
1 connector on the sync pulse generator.
25. Press the Measure button on the video measurement set to open the Measure
mode display.
26. Touch the Mode soft key to set the instrument to Analog mode, and then touch
the H_Timing soft key.
27. Press the Menu button to display the H_Timing main menu.
28. Touch the Average soft key and rotate the front-panel knob to set the value to
32.
29. Touch the RS-170A soft key.
30. Press the Select Line button.
31. Rotate the front-panel knob to set the measurement line to 100 (Field=1,
Line=100).
32. Verify that the burst amplitude is within the range of 271.4 mV to 286.4 mV.
33. Verify that the sync amplitude is within the range of 277.1 mV to 294.3 mV.
34. Move the BNC cable connection from CH 1 connector to the CH 2 connector
on the sync pulse generator.
35. Repeat steps 32 and 33.
36. Move the BNC cable connection from CH 2 connector to the CH 3 connector
on the sync pulse generator.
37. Repeat steps 32 and 33.
38. Move the BNC cable connection from CH 3 connector to the CH 4 connector
on the sync pulse generator.
39. Repeat steps 32 and 33.
SPG600 Option 02 Only. Perform steps 40 through 47 if Option 02 is installed in
your SPG600.
40. Move the BNC cable connection from CH 4 connector to the CH 5 connector
on the sync pulse generator.
41. Repeat steps 32 and 33.
42. Move the BNC cable connection from CH 5 connector to the CH 6 connector
on the sync pulse generator.
43. Repeat steps 32 and 33.
Performance Verification