Adjusting With External Weight (Only Kern Ew And Ewb) - KERN EG 220-3NM Operating Instruction

Precision balances ew-n series; eg-n series; ewb series
Hide thumbs Also See for EG 220-3NM:
Table of Contents

Advertisement

Turn the adjusting knob on the right side of the
balance to the WEIGH position.
Calibration is now finalised.
The balance will automatically return to weigh-
ing mode.

5.8.3 Adjusting with external weight (only KERN EW and EWB)

Carry out adjustment with the recommended adjusting weight (see Chapter 1 „Tech-
nical Data"). The adjustment can also be carried out with different adjusting weights
(see table), but not ideal from a metrological point of view. Information concerning
the adjusting weights is available at:
Model
EW 220-3NM
EW 420-3NM
EW 620-3NM
EW 820-2NM
EW 2200-2NM
EW 4200-2NM
EW 6200-2NM
EW 12000-1NM
EWB 220-2M
EWB 620-2M
EWB 1200-1M
EW/EG-N/EWB-e-1627
http://www.kern-sohn.com
Alternative adjusting weight
100 g
100 g
200 g
200 g
500 g
1000 g
2000 g
5000 g
100 g
200 g
500 g
CAL
buSY
End.
23