Introduction; Measurement Functions; Frequency A, B, C; Frequency Burst A, B, C - Fluke PM6690 Operator's Manual

Timer / counter / analyzer
Hide thumbs Also See for PM6690:
Table of Contents

Advertisement

Specifications

Introduction

Only values with tolerances or limits are guar-
anteed data. Values without tolerances are in-
formative data, without guarantee.
Measurement
Functions
Refer to page 8-9 for uncertainty information.
Inputs A and B can be swapped in all modes
except Rise Time and Fall Time.
Display: All measurements are displayed with
a large main parameter value and smaller aux-
iliary parameter values (with less resolution).
Some measurements are only available as aux-
iliary parameters.

Frequency A, B, C

n
Range:
Input A:
0.001 Hz - 300 MHz
Input B:
0.001 Hz - 300 MHz
Input C:
100 MHz - 3 GHz
(PM6690/6__)
200 MHz - 8 GHz
(PM6690/7__)
Resolution:
12 digits/s
n
Display:
Main Parameter:
Aux. Parameters:
8-2 Introduction
Frequency
V
, V
, V
max
min
p-p

Frequency Burst A, B, C

Frequency and PRF of repetitive burst signals
can be measured without external control sig-
nal and with selectable start arming delay.
Functions:
Frequency in burst (Hz)
PRF (Hz)
Number of cycles in burst
Range A, B, C:
See Frequency A, B, C
Min. Burst
Duration:
40 ns (80 ns > 160 MHz)
Min. No. of Pulses
in Burst (Inp A, B):
3 (6 above 160 MHz)
(Inp C):
3 x prescaler factor
PRF Range
(see also
Inp C spec):
0.5 Hz - 1 MHz
Start Delay
Range:
10 ns - 2 s, resolution
10 ns
n
Display:
Main Parameter:
Frequency in burst
Aux. Parameters:
PRF & number of cycles
in burst

Period A, B, C Average

n
Range:
Input A, B:
3.3 ns - 1000 s
Input C (3 GHz):
370 ps - 10 ns
(8 GHz):
125 ps - 5 ns
Resolution:
12 digits/s
n
Display:
Main Parameter:
Period
Aux. Parameters:
V
, V
, V
max
min
p-p

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents