The Telemetric Hr Signal Is Incorrect - Technogym Bike MED Service Maintenance Manual

Hide thumbs Also See for Bike MED:
Table of Contents

Advertisement

6.10.

THE TELEMETRIC HR SIGNAL IS INCORRECT

The machine displays this error if the receiver is disturbed by sources of electromagnetic noise.
Continued on the following page.
Page 6.20
THE TELEMETRIC
HR SIGNAL IS
INCORRECT
Check the earth connection
of the machine
Are the machines correctly
spaced by the required minimum
distance?
YES
Is the receiv er correctly
positioned and are the
connecting cables properly
routed?
YES
Are there sources of
electromagnetic noise near the
receiv er?
NO
A
BIKE MED: Service & Maintenance Manual - rev. 1.0
1
2
NO
Reposition the machines
so they are adequately
spaced
3
NO
Correct the position of the
receiv er and the routing of
the cables
4
Mov e the machine away
YES
f rom the noise source or
attenuate the power of the
noise

Hide quick links: