2.1.3 Recording Measurements

The EME Guard has a measurement recording function which can be activated or de-
activated with the help of the EME Guard Analysis V2.0 Software
The non-volatile internal memory enables a maximum number of 30,000 measurement
points (each measurement is time/date-stamped).
Measurement recording depends on the period and duration configured using the EME
Guard Analysis V2.0 Software.
This function enables the study of the level of exposure of the user to electric fields over a
predetermined period of time.
2.2
S
TARTING
Press the ON / OFF button.
The ON light is activated.
Maintain pressure for approximately one second until the ON light switches off and the
BATT. light comes on.
The device will then run an auto-test sequence enabling the user to check that all display
and alarm functions are operational.

2.2.1 Auto-test sequence of lights and alarms

The ON and BATT. lights are activated simultaneously for approximately 0.2 seconds.
The 25%, 50%, 75% and 100% lights are activated successively in this order then de-
activated in reverse order.
Each threshold reference light (20 V/m, 40 V/m, 60 V/m, 80 V/m, 100 V/m and 140
V/m lights in this order) is activated for approximately 0.1 seconds.
The audio alarm is activated for approximately 0.2 seconds.
The vibrating alarm is activated for approximately 0.2 seconds.
If one of these elements is not activated during the auto-test it is probably
defective.