Operation; Switching On/Off; Setting Instrument; Measurement Mode - TESTO 206 Instruction Manual

Hide thumbs Also See for 206:
Table of Contents

Advertisement

Available languages

Available languages

5. Operation

5.1 Switching on/off

Switch on instrument:
- All segments light up briefly and the instrument changes to the

measurement mode.

Switch off instrument: Keep

5.2 Setting instrument

The following functions can be set:
Function
Temperature unit
Auto Hold (AUTO HOLD)
Gradient / Offset
Calibration method (CAL)
Calibration points (CAL pH)
Auto Off (AUTO OFF)
Beeper (bP)
* Modification less than 0.02pH in 20s
The setting procedure can be interrupted by switching off the
instrument. The modifications are then not saved.
The instrument is switched off.
1 Opening the setting mode:
.
2 Select temperature unit (°C or °F):
Confirm selection:
3 Switch Hold on (On) or off (OFF):
Confirm selection:
- The gradient and offset values stored in the instrument are dis-
played as information.
4 Change view:
ON/HOLD
ON/HOLD
Description
Sets unit
Automatically holds reading,
as soon as it is stable*
Displays gradient and offset
values stored in the instrument
(amount)
Sets 1, 2 or 3 point
calibration
Sets calibration points
Instrument switches off
automatically after 10 minutes
if no button has been pressed
Warning sound (button pressed, On (switched on) or
stable reading reached with
Auto Hold function switched on)
MODE
MODE
.
MODE
.
.
MODE
.
pressed.
Setting options
°C or °F
On (switched on) or
OFF (switched off)
(Information only)
1P, 2P or 3P
1P: 4, 7 or 10
2P: 4 7 or 7 10
On (switched on) or
OFF (switched off)
OFF (switched off)
is kept pressed +
.
CAL
.
CAL
5. Operation
25
None
ON/HOLD