When A Shock Is Not Needed; Ecg/Cpr Dashboard - ZOLL AED 3 Operator's Manual

Automated external defibrillator
Hide thumbs Also See for AED 3:
Table of Contents

Advertisement

When a Shock is Not Needed

If the AED determines that a shock is not needed, a voice prompt
instructs you to start CPR for a period of time (configured by the AED
Administrator), after which the AED automatically initiates a new ECG
analysis. If a shock is needed at that time, refer to "When a Shock is
Needed" on page 16.
STEP 5
Start CPR
Place your hands on purple CPR compression sensor as shown on the
display screen. Listen to the metronome and follow the AED's
instructions. If the patient is an adult, follow the audio prompts that
provide feedback on the depth and the metronome rate for CPR chest
compressions. The CPR Countdown Timer displays the time left in the
current CPR period. Continue to follow the AED's voice instructions until
help arrives.

ECG/CPR Dashboard

If configured, the ZOLL AED 3 BLS model can display the CPR
Dashboard, with or without the ECG waveform. The CPR Dashboard
display shows the Chest Compression Rate, the Chest Compression
Depth, and the CPR remaining time until the next ECG analysis starts.
ZOLL AED 3 Operator's Manual
17

Advertisement

Table of Contents
loading

Table of Contents