Retracting Bands; Unable To Reach The Spool Retention Slots In The Guard Ports; Band Not Completely Unwound From Platform - ZOLL AutoPulse NXT User Manual

Table of Contents

Advertisement

Appendix B. Troubleshooting

Retracting bands

The band is installed. If the band immediately retracts when turning on the platform but without pressing
Start, do the following:
1. Turn off the platform.
2. Pull the band up to unwind the bands from the platform.
3. Remove the band guards and pins from the guard port. See "Removing the band" on page 19.
4. Install the band guards but not pins on the guard port. See "Installing the band" on page 16.
5. Turn on the platform.
The platform spools should return to the correct position. See Figure 4.2, "Platform spool," on page 17.
6. Remove the band guards.
7. Reinsert the band pins into the guard ports.
8. Reinstall the band guards.
9. Turn off the platform.
10. Turn on the platform.
The band should remain loose until the Start button is pressed.

Unable to reach the spool retention slots in the guard ports

If the spool in the guard port cannot be reached to install the band, do the following:
1. Install the band guards, but not the pins, in the guard port. See "Installing the band" on page 16.
2. Turn on the platform.
The platform spools should return to the correct position. See Figure 4.2, "Platform spool," on page 17.
3. Remove the band guards.
4. Insert the guard's pins into the guard ports.
5. Reinstall the band guards.

Band not completely unwound from platform

To manually remove the band from the platform, pull the band until the band is unwound and the pin is visible
and able to be removed.
Figure B.1. Band wrapped around platform spool
64
www.zoll.com

Advertisement

Table of Contents
loading