Levelflex FMP53 FOUNDATION Fieldbus
Severity of the event Default diagnostic category
Severity of the event Default diagnostic category
Low severity
1)
This event can be removed from the group in order to define its behavior individually; see section "Configurable
area".
Severity of the event Default diagnostic category
Lowest severity
Endress+Hauser
Out of specification (S)
Maintenance required (M)
Commissioning (block-based operation)
Source of the
Bit
event
Configuration
25
Process
24
Source of the
Bit
event
Sensor
23
Electronics
22
Configuration
21
Process
20
Source of the
Bit
event
Sensor
19
Elektronics
18
Configuration
17
Process
16
Events within the group
• C411: Up-/download
• C431: Trim
• C484: Simulation failure
mode
• C485: Simulation measured
value
• C491: Simulation current
output
• C585: Simulation distance
not used in Levelflex
Events within the group
not used in Levelflex
not used in Levelflex
S441: Current output 1
• S801: Energy too low
• S825: Operating temperature
• S921: Change of reference
• S942: In safety distance
• S943: In blocking distance
• S944: Level range
• S968: Level limited
Events within the group
not used in Levelflex
• M270: Main electronics
failure
• M272: Main electronics
failure
• M311: Electronics failure
M438: Data set
M803: Current loop 1
1)
47
Need help?
Do you have a question about the Levelflex FMP53 and is the answer not in the manual?