4.2.2 Water supply

Leica AutoStainer XL – Automated slide stainer
How to switch on:
1. Plug the power plug into the power socket.
2. Set the ON/STOP switch on the right side of the instrument to STOP.
3. Set the ON/OFF switch on the rear side of the instrument to ON.
4. Set the ON/STOP switch on the side to ON.
Then the instrument will emit 3 short beeps and the Main Menu will be
displayed.
When the instrument is not in use, set the ON/STOP switch on the side to
STOP.
The ON/OFF switch on the rear side of the instrument should be left ON.
The instrument must never be operated without the power
supply jumper cable.
The instrument must be connected to a grounded power
socket.
Connect the water hose to the water connection on the rear side of the in-
strument. Attach the other end of the hose to the cold-water tap. The hose
has a 3/4" connector. Then slowly turn the tap on all the way.
The water filter must be in place when the water hose is con-
nected, otherwise water can leak out.
Connect the drain hose to the discharge nozzle on the rear side of the
instrument.