5.
Operating the Instrument
5.4.2 Deleting program steps
5.4.3 Inserting a blank step into a program
26
Information contained in a step may be deleted, leaving a blank step.
1. Select the program
2. Select the step to be deleted
3. Press [F3] Erase.
Then the step is left blank.
You may enter new details for the step, if desired.
4. Save the program
This function is used to insert an additional step into an existing program.
1. Call up the program
2. Press [F2] Edit.
3. Select the step number where the new (blank) step is to be inserted.
4. Press [F1] More.
5. Press [F1] More.
6. Press [F1] Insert.
7. If you wish to proceed, press [F1] Yes.
A blank step is inserted now at the step selected under item 3.
The steps following the blank step will be renumbered.
Step 25 is lost when a blank step is inserted.
8. Continue editing the program.
9. Save the program
(see items 1 to 3 on page
(see item 4, page
(see page
27).
(see items 1 and 2 on page
(see page
27).
Instructions for Use V 2.5 – 01/2017
25).
25).
25).