AT235 Instructions for Use - US
3.5.5.1

Tymp test screen

A
Probe status showing the color corresponding to the probe light as
described in paragraph 3.3. It shows the labels: in ear, out of ear,
leaking or blocked
B
Probe tone frequency.
C
The current pressure is indicated in daPa.
D
The open triangle shows the current pressure. The solid triangle (in
manual mode (O) only) shows the target pressure.
E
The name of the current protocol.
F
Protocol list showing which test is currently viewed and in the check
boxes which tests will be tested after starting a test.
G
Press Prev. Test to select the previous test of the protocol list.
H
Press Next Test to select the next test of the protocol list.
I
Press Include to select or Exclude to deselect the checkbox of the
currently viewed test (F) and thereby include or exclude it from testing.
J
When several measurement attempts were done, pressing Data allows
choosing which data set is being viewed. Only the viewed data can be
saved to a client.