HP 8642A Operation And Calibration Manual page 287

Synthesized signal generator
Hide thumbs Also See for 8642A:
Table of Contents

Advertisement

Model 8642A/B
Performance Tests
PERFORMANCE TESTS
4-14. THIRD ORDER INTERMODULATION
SPECIFICATION:
Electrical Characteristics
Performance Limits
Conditions
Third Order Intermodulation
<-55 dBc.
+12 dBm. two generators 25
DESCRIPTION:
EQUIPMENT:
PROCEDURE:
kHz apart into a resistive
combiner.
Intermodulation is checked by summing another Signal Generator and the
HP 8642 into a spectrum analyzer. The intermodulation products are then
measured on the spectrum analyzer.
o
0
0
[
10
SPECTRUM
Rf
RF
0
0
0
~CJ~
§§ §§§§§
0
~
D
OUTPUT
INPUT
ANALYZER
I j
HP 8642A1B
SYNTHESIZED
>--
POWER SPL I TTER
StGNAL GENERATOR
6dB PAD
I
RF
OUTPUT
SIGNAL
GENERATOR
Figure
4 -1 O.
Intermodulation Test Setup
Spectrum Analyzer .......................... HP 8566B
Power Splitter ............................... HP
11667A
Signal Generator ............................ HP 8642B
6 dB Pad ..................................... HP
8491A
opt. 006
1.
Connect the instruments as shown in Figure 4-10.
2.
Set up the HP 8642 as follows:
INSTRUMENT PRESET
AMPLITUDE
FREQ
+12 dBm
1050 MHz
4-37

Advertisement

Table of Contents
loading

This manual is also suitable for:

8642b

Table of Contents