Operation
Model 8642A/B
Amplitude Sweep (cont'd)
3-88
Example
fOI'
Calculation #3
Set an ampli tude sweep from -10 dBm to - 21 dBrn wi th 100 amplit IIde steps.
Calculate the required sweep time setting for the
100 amplitude steps
using
calculation #
3:
Maximum number of steps is equal to:
1-10 dBm -
(-21
dBm)
I
0.1 dB
=
110
Minimum number of steps is equal
to a
minimum
of:
5
or
1-10 dBm -
(-21,
dBm)
I
0.1 dB
110
Since 5 is less than 110, the minimum number of steps is 5.
Therefore, It should be possible to select 100 steps since that value is between the
minimum value (5) and maximum value
(I
10).
To get 100 amplitllde steps, you can set the sweep time between the results of the
following calculatiol\S
(100
+
1)
X
3 ms
303 ms
(loa
+
2) X 3
ms
306 ms
SWEEP
SWEEP
SWEEP
DATA
~
AUTO
~
TI'vlE
OJwCD
.,
SII,RT
START
AMPTD
STAAT
c;[)
AMPTD
FPEO
mV
LOCAL
(keys)
SWEEP
DATA
SWEEP
DATA
SI,RI
8
CD w[][J
STOP
8
CIl
OJ
~
AMPTD
I\MPTJ
dem
~
(program
AASA ST 304 MS AA -10 DM AB -21 DM