How To Display The Increment Value - HP 8642A Operation And Calibration Manual

Synthesized signal generator
Hide thumbs Also See for 8642A:
Table of Contents

Advertisement

Operation
Model 8642A/B
Step, Increment Set
How to Display the Increment Value
HP-IB
Cocles
~~omments
3-150
To display the function's increment size, select the active function's key, then
I INCR SET
J.
Example
Set a 10 MHz increment set value for RF frequency. Then decrement the 100 MHz
RF frequency by 30 MHz.
LOCAL
(keys)
..
(program
codes)
Display
ENTRY
DATA
( FAEO)
(j~ff
I
OJ(]J
{
M~Z
]
FRIS 10 MZ FRDN FRDN FRDN
"T
-, f7I 171 f7I 171 f7I 171 f7I III
"7
, IL./. IL./ IL./ IL./ IL./ IL./ IL./ ,
j,
L
FREO
STEP
@] @] @]
, U
f7I f7I
T/
m'
.,
1 IL./.IL.I
.LI
A,
AMPrD
The increment value is displayed as it is entered. After the increment set value is
entered, the display returns to show the current instrument state. Note the change in
frequency value as the step key is pressed. (The digit the cursor is pointing to does not
change its value as the Step key is selected.
Refer to Knob, page 3-141, for
information about other ways to change values using the cursor positioning keys.)
Keys
Codes
INCR SET
IS'
Step Down
ON'
Step Up
UP'
"Function should be preceded by an active
function code.
Considerations When Using Increment Set with Amplitude Settings
If amplitude is displayed in dBm, and increment size is in volts, the amplitude setting
will change by the dB equivalent of the voltage increment.
Similarly, if amplitude is
displayed in volts and the increment setting is in dB, the amplitude setting will change
by the voltage-equivalent of the dB increment.
0.001 uV to 2.24V (0.002 EMF uV to 4.49 EMF V) increments can be selected.
However, the Signal Generator's setting resolution is limited to the voltage equivalent
of
0.1
dB at every output level (except during Linear Amplitude Sweep, Special
Function 121)