4.
Low Level SWR Setup
Connect the instruments as shown in Figure 4-8 step 2b.
Set the Signal Genera tor to:
INSTR UMENT PRESET
LEVEL:
FREQUENCY:
Set the HP 8642 to:
INSTRUMENT PRESET
FREQUENCY:
+16 dBm
4 MHz
500 MHz
Set up the measuring receiver as follows:
INSTRUMENT PRESET
MODE:
RF POWER
DISPLA Y MODE:
LOG
5. Low Level SWR, Reference
Set the Signal Generator and measuring receiver frequency to values
listed in the table below. Set the HP 8642 to the levels listed in the
table. Measure the "Ref." at all settings and record each reading
under "Intermediate Readings, Ref."
6.
Low Level SWR, Reflection
Connect the instruments as shown in Figure 4-8 step 2a. Put the
measuring receiver in RF POWER mode. Set the Signal Generator
and measuring receiver frequency to values listed in the table below.
Set the HP 8642 to the levels listed in the table. Measure the
reflected power at all settings and record each reading under
"Intermediate Readings, Reflected"
Signal
8642
Intermediate
Results
Generator
AMPTD
Readings
(Return loss)
Frequency
(dBm)
Ref.
Reflected
Actual
Min.
(MHz)
4.000
-14.0
14.0 dB
4.000
-24.0
14.0 dB
1000.000
-14.0
14.0 dB
1000.000
-24.0
14.0 dB
1300.000
-14.0
14.0 dB
1300.000
-24.0
14.0 dB