Operation
Model 8642A/B
Amplitude
(cont'd)
Procedures
Note:
Put the instrument in its initialized state by selecting [ INSTR PRESET 1.
3-70
AMPTD
How to Set the Amplitude Output (with the EMF Mode Off)
You can set the amplitude output by selecting the amplitude function and then setting
the desired output level:
I AMpTD I
and the desired output level (in dBm, V, mY, uV, or dBuV; the
I RAD %
J
key
is the dBuV terminator).
Example
Set the RF output to -107.3 dBm.
ENTRY
DATA
LOCAL
(keys)
[AMPfD )
8 OJ
IT]
OJ 0 OJ
[JD
..
(program
codes)
Display
I
f7I
f7I f7I f7jf7/ f7I f7I f7I
rn
7
I U
U.U U
U U U U III
L
FREQ
AP -107.3 OM
A'-'PTO
Amplitude has been set to -107.3 dBm. The cursor above the amplitude display setting
indicates amplitude is the active function.
Example
Set the RF output to 6 dBuV.
ENTRY
DATA
LOCAL
(keys)
!AMPm)
OJ
W
..
(program
AP 6 DU
codes)
Display
'V
5.0 ] ]
Ff;EQ
AMPTO
REl
Amplitude has been set to 6.0 dBuV. The AMPTD REL annunciators indicate that dBuV
is a relative setting. Even though dBuV can be set directly from the front panel, the way
this function operates and the way it is displayed are similar to most relative settings.
(Refer to Relative Amplitude, page 3-77.)
rev.20MAR85