HP 8642A Operation And Calibration Manual page 227

Synthesized signal generator
Hide thumbs Also See for 8642A:
Table of Contents

Advertisement

Model 8642A/B
Operation
Operating Messages (cont'd)
3) refer to the Sequence Dependency discussion that begins on page 3-12 in the HP-IB portion of this
Section III.
SP9 LIMITS MAX AMPTD
.E76
4076
Special Function 9 (Attenuator Range Hold) limits the maXImum amplitude that can be selected. To
resolve this problem, either
1)
select Attenuator Range Hold Off (Special Function 209), or
2)
select an amplitude within the held amplitude range selected (refer to Table 3-15, page 3--75, or Figure
3-9, page 3-76), or
3) refer to the Sequence Dependency discussion that begins on page 3-12 in the HP-IB portion of this
Section
III.
AMPTD+SP9 LIMIT MAX.AM
_E77
4077
The amplitude selected with Speci&l Function 9 (Attenuator Range Hold) limits the maximum value of
AM that can be selected. To resolve this problem, either
1)
select Attenuator Range Hold Off (Special Function 209), or
2)
reduce the amplitude setting to get the desired AM setting, or
3) select an AM setting allowed with the held range and amplitude setting selected (refer to Table 3-15,
page 3-75, or Figure 3-9, page 3-76), or
4)
refer to the Sequence Dependency discussion that begins on page 3-12 in the HP-IB portion of this
Section
III.
SP9 PREVENTS PULSE
.E78
4078
Pulse modulation is not possible when Special Function 9 (Attenuator Range Hold)
IS
selected. Select
Attenuator Range Hold Off (Special Function 209) before selecting pulse modulation .
PULSE PREVENTS SP9
. E79
4079
Special Function 9 (Attenuator Range Hold) is not possible when pulse modulation is selected. Turn off
pulse modulation (HP-IB: PLOF) and set output amplitude O-lP-IB: AP) to the desired level before
selecting attenuator range hold (HP-IB: SP9) .
SP9 PREVENTS SP4
. E80
4080
Special Function 4 (ALe Off) is not possible when Special Function 9 (Attenuator Range Hold) is selected.
Select Attenuator Range Hold Off (Special Function 209) before selecting ALe off.
HP-IB: Send SP209 before SP4.
SP4 PREVENTS SP9
.E81
4081
Special Function 9 (Attenuator Range Hold) is not possible when Special Function
4
(ALC Off) is selected.
Disable ALe Off (Special Function
204)
before selecting attenuator range hold.
HP-IB: Send SP204 before SP9.
AM PREVENTS SP4
.E82
4082
Special Function 4 (ALe Off) is not possible when AM is selected. Turn off AM before selecting ALe off.
HP-IB: Send AMOF before SP4.
PULSE PREVENTS SP4
.E83
4083
Special Function 4 (ALe Off) is not possible when pulse modulation
IS
selected.
Turn off pulse
modulation before selecting ALC off.
HP-IB: Send PLOF before SP4.
3-191

Advertisement

Table of Contents
loading

This manual is also suitable for:

8642b